Home

News

The current position:Home >> News
Precautions for use of X-ray machine
Time:2023-01-30 10:50:00 Number of hits:271

X-ray machine is a common equipment, so what should we pay attention to when using X-ray machine?

Before starting to use the equipment, first plug in the plug between the power supply and the host machine, and then plug in the network power supply plug. At the end, turn the other way around. First unplug the network power supply plug, and then unplug the plug between the power supply and the host machine: pay special attention to not press the button switch on the host machine when the power box is connected to the network power supply, otherwise the instrument may be seriously damaged.

When the equipment is not in use, it must be placed in an aluminum box. If it is not in use for the time being, it must be placed horizontally on the table to avoid damage. As long as it is not exposed to light, the light shield must be covered. Usually, when the aluminum box is stored, it should avoid falling down to avoid damage to the instrument.

The machine should be cleaned and maintained regularly. If the patient with blood needs to be scrubbed immediately after use, it can be scrubbed after screwing dry with alcohol. When scrubbing, water should not form water droplets to prevent water from seeping into the machine.

At ordinary times, pay attention not to allow foreign matters such as water, gypsum, alcohol, blood, etc. to enter the power switch, button or power plug to prevent circuit failure, which will affect the use.

If there is oil stain in the aluminum box during use, it may indicate that the ball tube has leaked oil, so it should be stopped immediately and contact the manufacturer immediately.

When the machine is used, the power socket must use a three-core socket with good grounding, otherwise it is easy to have high-voltage static electricity and may cause damage to the components. In order to prolong the service life of the machine, please avoid starting the machine continuously for a long time, with an interval of at least 5 seconds, so as to avoid that the previous high voltage does not completely retreat after the shutdown and the subsequent high voltage accumulates into an instantaneous high voltage, which breaks through the X-ray tube and causes artificial machine failure.

It should not be used by people who are sensitive to radiation. Children should use it with caution for eyes and genital parts.

The machine is relatively compact in structure and involves high voltage. Users are not allowed to disassemble it by themselves. If users find that the display is not on, the indicator light is not on, and the display is dark, please stop using it immediately and contact the manufacturer in time. So as not to cause failure.

If the image is found to have a moon-shaped defect during use, it generally indicates that the position of the ray part has shifted, which may lead to the occurrence of excessive radiation leakage. Stop using immediately and contact the manufacturer immediately.


If you have any questions, please contact us!

CONTACT US

LINK: linkedin twitter Facebook youtube

Copyright © 2022 Tianjin Wanji Network Technology Co., Ltd. All rights reserved   sitemaps  

Address:R2002, Building B, International Trading Mansion,Hexi District,Tianjin,China

Phone WhatsApp

Phone

+86 18630938527

Attention to us