Portable X-ray machine Portable X-ray machine Daily maintenance of portable X-ray machine
1. Daily maintenance
The X-ray machine should be placed in a clean and non-corrosive gas room. During the non-use period, it should be covered with dust-proof cloth or plastic cover, or placed in the box; The surface dirt needs to be cleaned with gauze dipped in detergent. Do not wipe the surface with alcohol or acid or alkaline reagent to avoid damaging the surface finish of the X-ray machine.
2. Maintenance during operation
The X-ray machine shall be operated in strict accordance with the requirements of the operation manual.
3. Maintenance during long-term parking
Parking environment
A. Ambient temperature: 10 ℃ - 40 ℃
B. Relative humidity: ≤ 30% - 75%
C. Atmospheric pressure: 700hPa - 1060hPa
4. Environmental requirements
Environmental precautions:
This machine is an X-ray diagnostic equipment. During the radiation output of the machine, the user should operate in the area behind the imaging, and can also wear effective protective equipment.
5. Disposal of environmental waste
After the machine has been used for a certain period of time, if the X-ray ball tube and transformer oil in the high-voltage generator need to be replaced, please send it back to our company or the local maintenance department designated by our company for replacement, and at the same time, the transformer oil should be recycled.